摘 要:采用6种标准物质测定CDR-1型差动热分析仪的差热分析法和差示扫描量热法
仪器常数,结果表明:CDR-1型差动热分析仪的差热分析法和差示扫描量热法仪器常数均随温
度的升高而增大,而差热分析法仪器常数受温度的影响更为显著。用最小二乘法得出在一定
条件下差热分析法和差示扫描量热法仪器常数与温度的函数关系。对KNO3样品的摩尔晶型转变
焓和摩尔熔化焓进行的对照测定表明:用相应转变温度下的仪器常数进行计算,比用单点标定
的仪器常数进行计算,可以较大程度地提高焓变测定的准确度。 |
| 关键词: |
|
文章全文为PDF格式,请下载到本机浏览。[下载全文]
如您没有PDF阅读器,请先下载PDF阅读器 Acrobat Reader [下载阅读器]
|
|
| Effect of temperature on the apparatus constant of |
|
|
| Abstract:The DTA and DSC apparatus constant of CDR-1 dynamic differential thermal analyzer was determined under different temperature by six kinds of standard matters. The result indicated that, DTA and DSC apparatus constants of CDR-1 dynamic differential thermal analyzer increase along with temperature increment,but influence of temperature on DTA apparatus constant is more remarkable. The functional relations of DTA and DSC apparatus constants for temperature were obtained by the least squares method under to a certainty condition. The balance determination forΔtΗm and ΔfΗm of KNO3 indicated that, the veracity of result, which was gained with apparatus constant under corresponding temperature, remarkably was enhanced. |